Disposable Temperature Probe

Classification: Imaging Diagnostic Equipment
Type: ECG
Certification: ISO13485
Weight: 0.05kg
Transport Package: Neutral
Specification: CE 13485
Customization:
Gold Member Since 2013

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company

Basic Info.

Model NO.
F1T5535
Trademark
F1
Origin
Made in China
HS Code
90181930
Production Capacity
10000

Product Description

For use with all YSI 400 series monitoring systems accepting

10K/2.252K/20K,50K/6.867K/15KOptional


Disposable Temperature Probe
1.0the usage range
  this specification suited for customer's appointed models.
2.0 usage condition(gb9706.1-2007 10.2.1)
2. 1: 37.5°rated temperature: 37.5°.
2.2: 25°~50°.
      operation environment: 10°~40°.
2.3.: ,storage and transport method: individual moistureproof package for each
      product, transport under the conditions of ambient temperature.
2.4: ac250v,: 2a.
        rated voltage: ac250v, current: 2a;
3.0: appearance requests:
a), , , the cable is round and glossy, the color uniformity, the jacket is ok, no
       disrepair, no dirt, no disfigurement, etc.
b), , , the surface of the connector is glossy, no burr, no obvious machine scar, no
      disfigurement, no modification, no dirt, no metal rust etc.
3.1(25°~50°).
      the probe end without shedding phenomenon(25°~50°).
3.2. product size in engineering drawing dimensioning
4.0technical requirements
4.1en12470. according to the standard of en12470
4.2. electrical performance:
4.2. 1 temperature accuracy test
4.2. 1. 1: 25°~50°. test conditions: 25°~50°.
4.2. 1. 2: 45°~50°, ±0.15°; 25°~45°, ±0.10°.
testing requirements: 45°~50°, accuracy is±0.15°of 25°~45°, accuracy is±0.10°.
4.3. mechanical properties:
4.3. 1±90° (ec53-1995 4.5. 5)
the connector should be done with 90 angle swing test. (ec53-1995 4.5. 5)
4.3. 2300g, ±90, 2500, 300mm,
  40/. (ec53-1995 4.5. 5)
main cable the swing test will weigh 300g, swing angle±90 degree,
2500 times, height 300mm. (ec53-1995 4.5. 5)
5.0standard
5.1rohs.
all the raw materials should comply with rohs standard
5.2iso 10993.
all of the processing of materials, products must meet iso 10993 standards.
6.0the accessory
6.1.
the proving documents of samples material.
6.2.
samples test report

Introduction:

We specialize in the design,development,manufacturing of the medical device and 

accessories,include ECG cable,EKG  cable,ECG electrodes,SPO2 sensor,Temperature 

probe,NIBP CUFF,etc.


Disposable Temperature ProbeDisposable Temperature ProbeDisposable Temperature ProbeDisposable Temperature ProbeDisposable Temperature Probe